ALTRUA
Report
- Report Number
- 2124215-2010-14597
- Event Type
- Malfunction
- Date Received
- October 5, 2010
- Date of Event
- July 15, 2010
- Report Date
- August 17, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED AND RESUBMITTED UPON RETURN AND COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT ONE DAY POST IMPLANT, THIS DEVICE EXHIBITED LOSS OF CAPTURE ON THE RIGHT ATRIAL (RA) CHANNEL. A REVISION PROCEDURE WAS PERFORMED. THE RA LEAD WAS REPOSITIONED AND TESTS OF THE LEAD CONFIRMED IT WAS OPERATING AS EXPECTED. AFTER RECONNECTING THE LEAD TO THE DEVICE, NOISE WAS NOTED WITH NO SENSING AND AN IMPEDANCE MEASUREMENT GREATER THAN 2500 OHMS. THE LEAD WAS DISCONNECTED, AND LEAD TESTING AGAIN NOTED NORMAL MEASUREMENTS. THE LEAD WAS RECONNECTED TO THE DEVICE SEVERAL TIMES, BUT THE ISSUE COULD NOT BE RESOLVED. THE DEVICE WAS REPLACED. NO ADVERSE PATIENT EFFECTS WERE OBSERVED.
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALTRUA | NVZ | GUIDANT CRM CLONMEL IRELAND | S208 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |