INRATIO
Report
- Report Number
- 2027969-2010-01483
- Event Type
- Malfunction
- Date Received
- September 27, 2010
- Date of Event
- August 26, 2010
- Report Date
- September 20, 2010
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K021923
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: (B)(6) 2010; INRATIO: 3.3; REFERENCE: 2.9; MEAN: 3.10; CONFIDENCE LIMITS: 1.8-4.2. DATE: (B)(6) 2010; INRATIO: 3.3; REFERENCE: 2.8; MEAN: 3.05; CONFIDENCE LIMITS: 1.8-4.2. THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND REFERENCE VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. INR REPORTED HAVE MET THE CRITERIA FOR ACCURACY. NO FURTHER INVESTIGATION WILL BE PURSUED. ANALYSIS OF THE CLIENT'S DATA FROM INRATIO AND REFERENCE TESTS REVEALED THAT TEST RESULT COMPARISON MET ACCURACY CRITERIA. CUSTOMER'S RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. STRIP LOT NUMBER WAS NOT PROVIDED BY CUSTOMER. THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING; CORRECTIVE ACTION NOT REQUIRED AT THIS TIME.
CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: INRATIO: 3.3; LAB: 2.9; COAGUCHECK: 2.8. TESTING WAS FIRST COMPLETED USING COAGUCHECK. ABOUT 2 MINUTES LATER TESTING WAS PERFORMED WITH THE INRATIO METER. ABOUT 10 MINUTES AFTER TESTING WITH INRATIO, THE LAB DRAW WAS TAKEN. PATIENT HAD JUST FINISHED TRAINING REVIEW.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | 100071 | NG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |