FDA Adverse Event Malfunction Summary report: N

FREESTYLE FREEDOM LITE

MDR report key: 1853175 · Received October 5, 2010

Report

Report Number
2954323-2010-01374
Event Type
Malfunction
Date Received
October 5, 2010
Date of Event
September 4, 2010
Report Date
October 4, 2010
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE METER HAS BEEN RETURNED AND INVESTIGATED. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. THE REPORTED READINGS WERE FOUND IN THE METER'S MEMORY LOG. THIS IS THE FINAL REPORT.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC METER. CUSTOMER REPORTED RECEIVING READINGS OF 94 MG/DL AND 453 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE FREEDOM LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 1013423

Patients

Seq Age Sex Outcome Treatment
1