FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 1853127 · Received October 5, 2010

Report

Report Number
6000001-2010-03738
Event Type
Malfunction
Date Received
October 5, 2010
Date of Event
September 9, 2010
Report Date
September 9, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF FAILURE CODE 18:115:1083:XXXX. THE ROOT CAUSE WAS DETERMINED TO BE A FAULTY USER INTERFACE MODULE PRINTED CIRCUIT BOARD. THE USER INTERFACE MODULE PRINTED CIRCUIT BOARD WAS REPLACED TO REPAIR THE REPORTED CONDITION. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.04.00. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE.

Description of Event or Problem · 1

THE BAXTER SERVICE TECHNICIAN REPORTED A COLLEAGUE INFUSION PUMP WHICH EXPERIENCED FAILURE CODE 18:115:1083:XXXX DURING DEVICE EVALUATION (SPECIFICALLY DURING ACCURACY TESTING). THERE WAS NO PATIENT INVOLVEMENT. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION FOR THIS DEVICE IS 5.04.00, CATEGORIZED AS UNREMEDIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1