FDA Adverse Event Malfunction Summary report: N

QUANTUM MAVERICK BALLOON CATHETER

MDR report key: 1853015 · Received October 5, 2010

Report

Report Number
2134265-2010-04354
Event Type
Malfunction
Date Received
October 5, 2010
Date of Event
September 4, 2010
Report Date
September 14, 2010
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT DEVICE WAS NOT RETURNED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS.(B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE A BALLOON RUPTURE OCCURRED. THE DETAILS AND LOCATION OF THE LESION ARE UNKNOWN. THE 3.5X15MM QUANTUM MAVERICK MONORAIL BALLOON RUPTURED. THE NUMBER OF INFLATIONS AND PRESSURE IS UNKNOWN. THE BALLOON WAS REMOVED INTACT. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUANTUM MAVERICK BALLOON CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493808015350 0012931998

Patients

Seq Age Sex Outcome Treatment
1