FREESTYLE NAVIGATOR
Report
- Report Number
- 2954323-2010-01371
- Event Type
- Malfunction
- Date Received
- October 5, 2010
- Date of Event
- September 17, 2010
- Report Date
- October 4, 2010
- Product Code
- LFR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THIS IS AN INITIAL REPORT. THE CUSTOMER'S PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. REMEDIAL ACTION (B)(4) WAS REPORTED TO THE (B)(4) OFFICE ON 14 APR 2009.
A CUSTOMER REPORTED THEY OBSERVED MOISTURE IN THE BATTERY COMPARTMENT OF THEIR FREESTYLE NAVIGATOR TRANSMITTER. IT HAS BEEN DETERMINED THAT SOME FREESTYLE NAVIGATOR TRANSMITTERS COULD POTENTIALLY EXHIBIT A CRACK/FRACTURE ON THE PLASTIC HOUSING NEAR THE BATTERY COMPARTMENT, WHICH COULD POTENTIALLY ALLOW MOISTURE TO COME IN CONTACT WITH THE TRANSMITTER'S INTERNAL ELECTRONICS, POTENTIALLY CAUSING THE TRANSMITTER AND THE RECEIVER TO LOSE CONNECTION INTERRUPTING THE CONTINUOUS GLUCOSE RESULTS. ALTHOUGH UNLIKELY, MOISTURE ENTERING THE TRANSMITTER HAS THE POTENTIAL TO GENERATE INACCURATE RESULTS ONLY WITH THE CONTINUOUS GLUCOSE READINGS. THE BUILT-IN FREESTYLE GLUCOSE METER IS NOT IMPACTED BY THIS ISSUE AND THE USER CAN CONTINUE TO USE THE BUILT-IN METER. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.