FDA Adverse Event Injury Summary report: N

FLEXTEND II

MDR report key: 1852637 · Received October 4, 2010

Report

Report Number
2124215-2010-14357
Event Type
Injury
Date Received
October 4, 2010
Date of Event
July 14, 2010
Report Date
August 27, 2010
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

BOSTON SCIENTIFIC, CRM WILL ANALYZE THIS DEVICE UPON RECEIPT AND PROVIDE ADDITIONAL INFORMATION ONCE THE TESTING HAS BEEN COMPLETED AND AN UPDATED REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE COMPLETE LEAD WAS PERFORMED. VISUALLY, THE LEAD WAS RETURNED WITH THE HELIX RETRACTED AND DRIED BLOOD WAS OBSERVED IN THE LUMEN AND DRIED TISSUE WAS FOUND ON THE HELIX. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY; MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. ELECTRICALLY, THE LEAD WAS FOUND TO BE WITHIN SPECIFICATION AND THE ALLEGATION OF DISLODGEMENT COULD NOT BE CONFIRMED BY TESTING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING IMPLANT, THIS RIGHT VENTRICULAR (RV) PACE/SENSE LEAD EXHIBITED HIGH PACING THRESHOLDS (< 4.5V AT 0.5MS) SO MULTIPLE ATTEMPTS WERE MADE TO REPOSITION THE LEAD. DESPITE INCREASED THRESHOLDS, THE RIGHT ATRIAL LEAD WAS IMPLANTED AND THE POCKET WAS CLOSED. ANOTHER MEASUREMENT SHOWED CONTINUED, INCREASED THRESHOLDS AND DECREASED SENSING SO THE POCKET WAS REOPENED TO REPOSITION THE RV LEAD; DESPITE ADDITIONAL ATTEMPTS TO REPOSITION THE LEAD, IT CONTINUED TO DISLODGE SO THE RV LEAD WAS FINALLY EXPLANTED. UPON EXPLANT, THE PHYSICIAN NOTED DIFFICULTY WHEN THEY TRIED TO EXTEND THE HELIX USING THE FIXATION TOOL SO THE RV LEAD WAS REPLACED SUCCESSFULLY. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXTEND II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4097

Patients

Seq Age Sex Outcome Treatment
1