FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 1852601 · Received October 4, 2010

Report

Report Number
6000001-2010-03726
Event Type
Malfunction
Date Received
October 4, 2010
Date of Event
September 14, 2010
Report Date
September 14, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 808:02 WAS DISCOVERED AND CONFIRMED DURING PRODUCT EVALUATION IN THE PUMP'S EVENT HISTORY. THIS CONDITION WAS CAUSED BY A DEFECTIVE PUMPHEAD MODULE. THE PUMPHEAD MODULE WAS REPLACED TO CORRECT THE REPORTED CONDITION. ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

DURING SERVICE BY A BAXTER SERVICE TECHNICIAN, A COLLEAGUE INFUSION PUMP WAS FOUND WITH FAILURE CODE 808:02 ON CHANNEL A. THIS CONDITION INTERRUPTED DELIVERY AT THE START OF THE ACCURACY TEST. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.04.00.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1