FDA Adverse Event Malfunction Summary report: N

BIOFIRE® BLOOD CULTURE IDENTIFICATION 2 (BCID2) PANEL

MDR report key: 18525084 · Received January 17, 2024

Report

Report Number
3002773840-2024-00179
Event Type
Malfunction
Date Received
January 17, 2024
Manufacturer
BIOFIRE DIAGNOSTICS, LLC
Product Code
PAM
UDI-DI
00815381020338
PMA / PMN Number
K193519
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: A POTENTIAL FALSE POSITIVE C. TROPICALIS RESULT ON THE BIOFIRE BCID2 PANEL WAS REPORTED AFTER TESTING A PATIENT BLOOD CULTURE SAMPLE ON AN UNKNOWN DATE. IT IS UNKNOWN IF THE PATIENT WAS IMPACTED DUE TO THE BIOFIRE BCID2 RESULT. NO SERIOUS INJURY OR DEATH OCCURRED. CONCLUSION: THE INVESTIGATION DETERMINED THAT THE MOST LIKELY CAUSE OF THE DISCREPANT C. TROPICALIS RESULT WAS THE PRESENCE OF NON-VIABLE ORGANISM/NUCLEIC ACID IN THE BLOOD CULTURE MEDIA BOTTLE. SIMILAR EVENTS WERE RECENTLY OBSERVED IN THE FIELD WITH THE COMBINATION OF THE BIOFIRE BCID2 PANEL AND BD BACTEC¿ BLOOD CULTURE BOTTLES. CUSTOMERS WERE NOTIFIED OF THIS ISSUE VIA CUSTOMER SERVICE NOTIFICATION REFERENCE#: CSN-2023-290. WHILE BLOOD CULTURE VIALS ARE AUTOCLAVED AND ROUTINELY QUALITY CONTROLLED FOR STERILITY, NON-VIABLE ORGANISMS OR NUCLEIC ACIDS CAN REMAIN IN THE BLOOD CULTURE MEDIA AFTER THE STERILIZATION PROCESS. THE PRESENCE OF NON-VIABLE ORGANISMS AND NUCLEIC ACID DOES NOT COMPROMISE THE INTENDED USE OF BLOOD CULTURE MEDIA, CULTURING VIABLE MICROORGANISMS; HOWEVER, THE BIOFIRE BCID2 PANEL DOES NOT DISTINGUISH BETWEEN NUCLEIC ACID FROM VIABLE OR NON-VIABLE ORGANISMS. THE BIOFIRE BCID2 PANEL IS WORKING AS INTENDED. THE "LABORATORY PRECAUTION" AND "LIMITATION" SECTIONS OF THE BIOFIRE BCID2 PANEL INSTRUCTIONS FOR USE (IFU) (WWW.ONLINE-IFU.COM/ITI0048) OUTLINES THE POTENTIAL FOR FALSE POSITIVE DETECTIONS DURING MOLECULAR TESTING WITH STERILE BLOOD CULTURE MEDIA CONTAINING DETECTABLE LEVELS OF NON-VIABLE ORGANISMS AND/OR NUCLEIC ACID. IMPORTANTLY, RESULTS FROM THE BIOFIRE BCID2 PANEL ARE INTENDED TO BE CORRELATED WITH THE CLINICAL HISTORY, EPIDEMIOLOGICAL DATA, AND OTHER DATA AVAILABLE TO THE CLINICIAN EVALUATING THE PATIENT. ALL IDENTIFICATION RESULTS PROVIDED BY THE BIOFIRE BCID2 PANEL ARE INTENDED TO BE INTERPRETED IN CONJUNCTION WITH GRAM STAIN RESULTS. CLINICAL PERFORMANCE: ACCORDING TO TABLE 38. BIOFIRE BCID2 PANEL CLINICAL PERFORMANCE SUMMARY, CANDIDA SPP. OF THE BIOFIRE BCID2 PANEL IFU, THE PERFORMANCE CLAIM FOR THE C. TROPICALIS ASSAY COMPARED TO STANDARD OF CARE IDENTIFICATION FOR GENUS LEVEL FOLLOWED BY PCR & SEQUENCING OF ISOLATES FOR SPECIES IDENTIFICATION SHOWED AN OVERALL SENSITIVITY OF 100% (95% CI 93.5-100%) AND AN OVERALL SPECIFICITY OF 99.9% (95% CI 99.7-100%). THE SINGLE FALSE POSITIVE SPECIMEN WAS IDENTIFIED AS A CROSS-REACTIVITY BETWEEN THE BIOFIRE BCID2 PANEL C. TROPICALIS ASSAY AND HIGH TITER CANDIDA PARAPSILOSIS. THIS CROSS-REACTIVITY IS A KNOWN LIMITATION OF THE BIOFIRE BCID2 PANEL.

Description of Event or Problem · 0

SUMMARY: YUMA REGIONAL MEDICAL CENTER (US) REPORTED A POTENTIAL FALSE POSITIVE CANDIDA TROPICALIS RESULT ON THE BIOFIRE BLOOD CULTURE IDENTIFICATION 2 (BCID2) PANEL AFTER TESTING A PATIENT BLOOD CULTURE SAMPLE. DUE TO THE BIOFIRE BCID2 PANEL RESULT, IT IS UNKNOWN IF THE PATIENT WAS HARMED OR AFFECTED. THE INVESTIGATION DETERMINED THAT THE MOST LIKELY CAUSE OF THE DISCREPANT C. TROPICALIS RESULT WAS THE PRESENCE OF NON-VIABLE ORGANISM/NUCLEIC ACID IN THE BLOOD CULTURE MEDIA BOTTLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1845015 BIOFIRE® BLOOD CULTURE IDENTIFICATION 2 (BCID2) PANEL BIOFIRE® BLOOD CULTURE IDENTIFICATION 2 (BCID2) PANEL PAM BIOFIRE DIAGNOSTICS, LLC RFIT-ASY-0147 00815381020338

Patients

Seq Age Sex Outcome Treatment
1 Unknown