FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 1852193 · Received September 29, 2010

Report

Report Number
1720753-2010-03268
Event Type
Malfunction
Date Received
September 29, 2010
Date of Event
September 16, 2010
Report Date
September 29, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE HIGH VOLTAGE CABLE AND THE FLUORO FUNCTIONS BOARD WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 9800 SYSTEM HAD A COLLIMATOR IRIS ERROR DURING A CASE. THE 9800 SYSTEM HAD TO BE REBOOTED AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1