FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 1852171 · Received October 4, 2010

Report

Report Number
2124215-2010-18251
Event Type
Malfunction
Date Received
October 4, 2010
Date of Event
July 12, 2010
Report Date
July 12, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE. THIS ISSUE IS DISCUSSED IN THE Q3 2010 PRODUCT PERFORMANCE REPORT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), WHICH IS PART OF THE MID-LIFE DISPLAY OF REPLACEMENT INDICATORS PRODUCT UPDATE CLASSIFIED AS A PRODUCT ADVISORY AND INITIALLY COMMUNICATED ON 3/10/2007, DISPLAYED A BATTERY STATUS OF ELECTIVE REPLACEMENT INDICATOR (ERI). THE MONITORING VOLTAGE WAS 2.69 VOLTS AND THE CHARGE TIME WAS 22.8 SECONDS. THERE WAS A CONCERN THAT THE BATTERY DEPLETED EARLIER THAN EXPECTED SINCE A MONTH PRIOR, THE CHARGE TIME MEASUREMENT WAS 16.9 SECONDS. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS CLINICAL OBSERVATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T167

Patients

Seq Age Sex Outcome Treatment
1 63 YR 4470| T167| 0185