VITALITY 2
Report
- Report Number
- 2124215-2010-18251
- Event Type
- Malfunction
- Date Received
- October 4, 2010
- Date of Event
- July 12, 2010
- Report Date
- July 12, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE. THIS ISSUE IS DISCUSSED IN THE Q3 2010 PRODUCT PERFORMANCE REPORT.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), WHICH IS PART OF THE MID-LIFE DISPLAY OF REPLACEMENT INDICATORS PRODUCT UPDATE CLASSIFIED AS A PRODUCT ADVISORY AND INITIALLY COMMUNICATED ON 3/10/2007, DISPLAYED A BATTERY STATUS OF ELECTIVE REPLACEMENT INDICATOR (ERI). THE MONITORING VOLTAGE WAS 2.69 VOLTS AND THE CHARGE TIME WAS 22.8 SECONDS. THERE WAS A CONCERN THAT THE BATTERY DEPLETED EARLIER THAN EXPECTED SINCE A MONTH PRIOR, THE CHARGE TIME MEASUREMENT WAS 16.9 SECONDS. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS CLINICAL OBSERVATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T167 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | 4470| T167| 0185 |