I-STAT CG4+ CARTRIDGE
Report
- Report Number
- 2245578-2024-00010
- Event Type
- Malfunction
- Date Received
- January 16, 2024
- Date of Event
- December 31, 2023
- Report Date
- February 9, 2024
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- KHP
- UDI-DI
- 10054749000132
- PMA / PMN Number
- K982071
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 31-JAN-2024. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AL (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED FOR CG4+ CARTRIDGE LOT M23227.
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON 02-JAN-2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CG4+ CARTRIDGE THAT YIELDED A SUSPECTED DISCREPANT PO2 RESULT ON A 185 MMHG ON A 63 YEAR OLD MALE PATIENT WITH END-STAGE CHRONIC RENAL FAILURE ON REGULAR DIALYSIS, TYPE 2 DIABETES, HYPERTENSION, PNEUMONIA , LIPID DISORDERS, AND GASTROESOPHAGEAL REFLUX. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE TESTED PO2 SO2 SAMPLE I-STAT (B)(6) 223 0949 25 MMHG 40% A I-STAT (B)(6) 223 NI 185 MMHG 99.6 B THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 854439 | I-STAT CG4+ CARTRIDGE | CG4+ CARTRIDGE | KHP | ABBOTT POINT OF CARE | NA | M23227 | 10054749000132 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Male |