FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 1852068
·
Received September 28, 2010
Report
- Report Number
- 1720753-2010-03246
- Event Type
- Malfunction
- Date Received
- September 28, 2010
- Date of Event
- August 2, 2010
- Report Date
- September 28, 2010
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE TUBE WAS CALIBRATED AND THE SOFTWARE INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 9800 SYSTEM HAD A HIGH MA ERROR DURING A CASE. THE 9800 SYSTEM HAD TO BE REBOOTED AND THE PROCEDURE WAS COMPLETED WITHOUT INCIDENT. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |