FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 1852055
·
Received September 28, 2010
Report
- Report Number
- 1720753-2010-03248
- Event Type
- Malfunction
- Date Received
- September 28, 2010
- Date of Event
- September 8, 2010
- Report Date
- September 28, 2010
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE CABLE WAS REPAIRED AND THE IMAGE PROCESSOR, VIDEO CONTROLLER, AND DISPLAY ADAPTOR WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 9800 SYSTEM HAD POOR IMAGE QUALITY WITH TOO DARK IMAGES AND A LINE IN THE MIDDLE OF THE IMAGE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |