FDA Adverse Event Malfunction Summary report: N

2800

MDR report key: 1852049 · Received September 28, 2010

Report

Report Number
1720753-2010-03235
Event Type
Malfunction
Date Received
September 28, 2010
Date of Event
September 15, 2010
Report Date
September 28, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE PS3 POWER SUPPLY WAS ADJUSTED. THE SYSTEM INTERFACE BOARD, SINGLE BOARD COMPUTER, HARD DRIVE, AND THE SOFTWARE UPGRADE WERE INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED, AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 2800 SYSTEM HAD THE COLLIMATOR CLOSE DOWN DURING A CASE. THE PROCEDURE WAS COMPLETED WITHOUT INCIDENT. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 2800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 2800

Patients

Seq Age Sex Outcome Treatment
1