FDA Adverse Event
Malfunction
Summary report: N
9600
MDR report key: 1852036
·
Received September 28, 2010
Report
- Report Number
- 1720753-2010-03240
- Event Type
- Malfunction
- Date Received
- September 28, 2010
- Date of Event
- September 14, 2010
- Report Date
- September 28, 2010
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE CAMERA, ETHERNET CABLE, AND DICOM BOX WERE REPLACED AND THE 5 VOLT POWER SUPPLY WAS ADJUSTED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 9600 SYSTEM LOCKED UP AND WOULD NOT SAVE IMAGES OR SEND TO PACS. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9600 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |