FDA Adverse Event Malfunction Summary report: N

PUMP MMT-511LNAS PRDGM INS SK EN US LN

MDR report key: 1852016 · Received September 24, 2010

Report

Report Number
2032227-2010-82751
Event Type
Malfunction
Date Received
September 24, 2010
Date of Event
September 5, 2010
Report Date
September 5, 2010
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
K001829
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A MOTOR ERROR ALARM DURING THE MANUAL PRIME. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-511LNAS PRDGM INS SK EN US LN INFUSION PUMP LZG MEDTRONIC MINIMED MMT-511LNAS

Patients

Seq Age Sex Outcome Treatment
1 30 YR