FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-511LNAS PRDGM INS SK EN US LN
MDR report key: 1852016
·
Received September 24, 2010
Report
- Report Number
- 2032227-2010-82751
- Event Type
- Malfunction
- Date Received
- September 24, 2010
- Date of Event
- September 5, 2010
- Report Date
- September 5, 2010
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- LZG
- PMA / PMN Number
- K001829
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A MOTOR ERROR ALARM DURING THE MANUAL PRIME. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-511LNAS PRDGM INS SK EN US LN | INFUSION PUMP | LZG | MEDTRONIC MINIMED | MMT-511LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 YR |