FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE-TREAT SURFACE&SCALE
MDR report key: 1851974
·
Received September 24, 2010
Report
- Report Number
- 1824206-2010-10099
- Event Type
- Malfunction
- Date Received
- September 24, 2010
- Date of Event
- August 30, 2010
- Report Date
- August 30, 2010
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN FOUND THE HEAD UP VALVE IS STUCK. THE CAUSE IS CONTAMINATION IN THE HYDRAULIC FLUID. TECHNICIAN REPLACED THE HEAD UP VALVE TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
TECHNICIAN ALLEGED THAT THE HEAD WILL NOT GO UP. NO ALLEGED INJURIES BY ACCOUNT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE-TREAT SURFACE&SCALE | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | P1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|