FDA Adverse Event Malfunction Summary report: N

TOTALCARE-TREAT SURFACE&SCALE

MDR report key: 1851974 · Received September 24, 2010

Report

Report Number
1824206-2010-10099
Event Type
Malfunction
Date Received
September 24, 2010
Date of Event
August 30, 2010
Report Date
August 30, 2010
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND THE HEAD UP VALVE IS STUCK. THE CAUSE IS CONTAMINATION IN THE HYDRAULIC FLUID. TECHNICIAN REPLACED THE HEAD UP VALVE TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

TECHNICIAN ALLEGED THAT THE HEAD WILL NOT GO UP. NO ALLEGED INJURIES BY ACCOUNT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE-TREAT SURFACE&SCALE A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. P1900

Patients

Seq Age Sex Outcome Treatment
1