FDA Adverse Event Malfunction Summary report: N

COMMUNICATOR

MDR report key: 1851944 · Received October 4, 2010

Report

Report Number
2124215-2010-14580
Event Type
Malfunction
Date Received
October 4, 2010
Date of Event
June 30, 2010
Report Date
July 8, 2010
Manufacturer
EXTERNAL MANUFACTURER
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMMUNICATOR ALONG WITH THE POWER SUPPLY WERE RETURNED FOR ANALYSIS. POST MARKET QUALITY ASSURANCE FOUND THAT THE RETURNED POWER BRICK GOT HOT TO THE TOUCH. ANALYSIS ALSO CONFIRMED THAT THE PLASTIC HOUSING WAS DEFORMED, AND HAD A BURNING SMELL.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION FROM THE PATIENT THAT THE COMMUNICATOR HAD SCREEN FAILURE AFTER A RECENT STORM. A REPLACEMENT COMMUNICATOR WAS SHIPPED TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMMUNICATOR ACCESSORY LWS EXTERNAL MANUFACTURER 6482

Patients

Seq Age Sex Outcome Treatment
1 75 YR 4542| H217| 4086| 0158