FDA Adverse Event
Malfunction
Summary report: N
COMMUNICATOR
MDR report key: 1851944
·
Received October 4, 2010
Report
- Report Number
- 2124215-2010-14580
- Event Type
- Malfunction
- Date Received
- October 4, 2010
- Date of Event
- June 30, 2010
- Report Date
- July 8, 2010
- Manufacturer
- EXTERNAL MANUFACTURER
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COMMUNICATOR ALONG WITH THE POWER SUPPLY WERE RETURNED FOR ANALYSIS. POST MARKET QUALITY ASSURANCE FOUND THAT THE RETURNED POWER BRICK GOT HOT TO THE TOUCH. ANALYSIS ALSO CONFIRMED THAT THE PLASTIC HOUSING WAS DEFORMED, AND HAD A BURNING SMELL.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION FROM THE PATIENT THAT THE COMMUNICATOR HAD SCREEN FAILURE AFTER A RECENT STORM. A REPLACEMENT COMMUNICATOR WAS SHIPPED TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COMMUNICATOR | ACCESSORY | LWS | EXTERNAL MANUFACTURER | 6482 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | 4542| H217| 4086| 0158 |