FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 1851892 · Received October 4, 2010

Report

Report Number
2124215-2010-14100
Event Type
Malfunction
Date Received
October 4, 2010
Date of Event
June 28, 2010
Report Date
July 15, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT, THE DEVICE WAS SUCCESSFULLY INTERROGATED BY BOSTON SCIENTIFIC'S POST MARKET QUALITY ASSURANCE LABORATORY. ENGINEERING CALCULATIONS DETERMINED THAT THIS DEVICE DID NOT MEET EXPECTED LONGEVITY. THE DEVICE IS CURRENTLY UNDERGOING LABORATORY TESTING TO DETERMINE ROOT CAUSE.

Additional Manufacturer Narrative · 1

THE DEVICE WAS PUT THROUGH AND PASSED A SERIES OF AUTOMATED TESTS WHICH VERIFIED FUNCTIONALITY, AND THERAPY DELIVERY. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. ALL LOW VOLTAGE POWER SUPPLY MEASUREMENTS WERE CURRENTLY WITHIN AN APPROPRIATE RANGE; HOWEVER, FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION. THIS ISSUE IS DISCUSSED IN THE Q2 2010 PRODUCT PERFORMANCE REPORT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS DEVICE, IMPLANTED IN (B)(6) 2006, TRIGGERED ELECTIVE REPLACEMENT INDICATOR (ERI) IN (B)(6) 2010. THERE WERE NO ALLEGATIONS OR COMPLAINTS AGAINST THE DEVICE'S OPERATION, FUNCTIONALITY OR LONGEVITY.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T167

Patients

Seq Age Sex Outcome Treatment
1 50 YR 4470| T167| 4480| E110| 0175