FDA Adverse Event Malfunction Summary report: N

2.8MM THRDD GUIDE F/ 3.5MM SCR VA AND LCP

MDR report key: 18518855 · Received January 16, 2024

Report

Report Number
8030965-2024-00897
Event Type
Malfunction
Date Received
January 16, 2024
Date of Event
December 19, 2023
Manufacturer
SYNTHES GMBH
Product Code
FZX
UDI-DI
10886982269973
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H6: A PRODUCT INVESTIGATION WAS CONDUCTED. THE DEVICE ASSOCIATED WITH THIS REPORT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. VISUAL ANALYSIS AND INSPECTION OF THE DEVICE FOUND NO DAMAGE OR DEFECT FOUND. A FUNCTIONAL TEST CANNOT BE PERFORMED WITHOUT MATING DEVICE, THEREFORE IT IS NOT POSSIBLE TO CONFIRM THE REPORTED ISSUE. A DIMENSIONAL INSPECTION WAS PERFORMED AND MET SPECIFICATIONS. THE OVERALL COMPLAINT WAS UNCONFIRMED AS THE OBSERVED CONDITION OF THE THREADED DRILL GUIDE 2.8 F/SCREW 3.5 F/V WOULD NOT CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. H3, H4, H6: DEVICE HISTORY RECORD (DHR) REVIEW CONDUCTED: PART: 03.133.004 SYNTHES LOT: J012291 SUPPLIER LOT: J012291 RELEASE TO WAREHOUSE DATE: 02 MAR 2023 SUPPLIER: (B)(4), NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D9: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. H3, H6: THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE PATIENT UNDERWENT A ORIF FOR DISTAL MEDIAL TIBIA & DISTAL ANTEROLATERAL TIBIA. ADT PLATE WAS USED ON DISTAL ANTEROLATERAL TIBIA, AND 1/3 CIRCLE PLATE WAS USED ON DISTAL MEDIAL TIBIA. DURING THE SURGERY, A DRILL-GUIDE/3.5MM WITH A SCREW (03.133.004) WAS NOT ATTACHED TO THE PLATE. THEREFORE, INSTEAD OF LOCKING SCREW, CORTEX ONE WAS INSERTED IN THE PLATE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 2.8MM THRDD GUIDE F/ 3.5MM SCR VA AND LCP. THIS IS REPORT 3 OF 5 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1627832 2.8MM THRDD GUIDE F/ 3.5MM SCR VA AND LCP GUIDE, SURGICAL, INSTRUMENT FZX SYNTHES GMBH J012291 10886982269973

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown LCP ONE-THIRD TUBULAR PLATE 3.5, 5 HOLES| THREADED DRILL GUIDE 2.8 F/SCREW 3.5 F/V| THREADED DRILL GUIDE 2.8 F/SCREW 3.5 F/V| THREADED DRILL GUIDE 2.8 F/SCREW 3.5 F/V