STARCLOSE SE VASCULAR CLOSURE SYSTEM
Report
- Report Number
- 2953144-2010-02224
- Event Type
- Injury
- Date Received
- September 29, 2010
- Date of Event
- December 10, 2009
- Report Date
- December 11, 2009
- Manufacturer
- ABBOTT VASCULAR-VASCULAR SOLUTIONS
- Product Code
- MGB
- PMA / PMN Number
- P050007
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS PROVIDED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT. ADDITIONALLY, 3 STERILE UNUSED DEVICES FROM THE SAME LOT WERE RECEIVED AND WERE FUNCTIONALLY TESTED. THE DEVICES PASSED FUNCTIONAL TESTING WITH ACCEPTABLE RESULTS. NO MFG OR QUALITY ISSUE WAS DETECTED.
IT WAS REPORTED THAT THE PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE SE DEVICE ACHIEVED ARTERIOTOMY CLOSURE OF AN UNSPECIFIED VESSEL AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, AFTER CLIP DEPLOYMENT THE DEVICE WAS DIFFICULT TO REMOVE FROM THE PT'S ARTERY. IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE, AN ATTEMPT WAS MADE TO USE THE ACCESS PORTS, HOWEVER, IT WAS UNSUCCESSFUL. THE DEVICE WAS ULTIMATELY REMOVED USING COUNTER-TRACTION AND ASSERTIVE PULL. HEMOSTASIS WAS ACHIEVED WITH THE CLIP. THERE WERE NO REPORTED ADVERSE PT EFFECTS. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STARCLOSE SE VASCULAR CLOSURE SYSTEM | MGB | ABBOTT VASCULAR-VASCULAR SOLUTIONS | NA | 81032-6H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |