FDA Adverse Event Injury Summary report: N

STARCLOSE SE VASCULAR CLOSURE SYSTEM

MDR report key: 1851751 · Received September 29, 2010

Report

Report Number
2953144-2010-02245
Event Type
Injury
Date Received
September 29, 2010
Date of Event
January 28, 2010
Report Date
February 5, 2010
Manufacturer
ABBOTT VASCULAR - VASCULAR SOLUTIONS
Product Code
MGB
PMA / PMN Number
P050007
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS NOT PROVIDED; THEREFORE, A DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE SE DEVICE ACHIEVED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, AFTER DEPLOYING THE CLIP, THE DEVICE WAS DIFFICULT TO REMOVE. IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE, A DILATOR WAS INSERTED INTO THE ACCESS PORTS UNLOCKING THE THUMB ADVANCER. THE DEVICE RELEASED FROM THE TISSUE TRACT AND DEPLOYED THE CLIP. THE CLIP FROM THE STARCLOSE SE DEVICE ACHIEVED HEMOSTASIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STARCLOSE SE VASCULAR CLOSURE SYSTEM MGB ABBOTT VASCULAR - VASCULAR SOLUTIONS NA UNK

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention