ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2010-14235
- Event Type
- Injury
- Date Received
- October 4, 2010
- Date of Event
- July 12, 2010
- Report Date
- July 20, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
AN ATTEMPT HAS BEEN MADE TO GATHER ADDITIONAL INFORMATION ABOUT THIS EVENT. THIS EVENT WILL BE UPDATED IF ANY ADDITIONAL INFORMATION IS RECEIVED.
ADDITIONAL INFORMATION INDICATES THAT THE INCREASE IN RV PACING THRESHOLDS DID LEAD TO SOME RV LOSS OF CAPTURE (LOC). SUBSEQUENTLY, A LEAD REVISION WAS PERFORMED, DURING WHICH THIS RV LEAD WAS SUCCESSFULLY REPOSITIONED WITHOUT ANY ADVERSE PATIENT EFFECTS RESULTING. THE POSSIBILITY OF AN RV LEAD MICRODISLODGEMENT OR A CHANGE IN PATIENT CONDITION WAS DISCUSSED. AVAILABLE INFORMATION SUGGESTS THAT THIS RV LEAD REMAINS IN SERVICE AT THIS TIME. THIS EVENT WILL BE UPDATED IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RECENTLY IMPLANTED TRANSVENOUS RIGHT VENTRICULAR (RV) LEAD WAS EXHIBITING INCREASED PACING THRESHOLDS OF 4.5 V, AND A DROP IN R-WAVE SENSING FROM 25 MV TO 9 MV. LEAD IMPEDANCE MEASUREMENTS REMAIN WITHIN NORMAL RANGE. THE POSSIBILITY OF AN RV LEAD DISLODGEMENT WAS DISCUSSED. NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED AS A RESULT OF THESE OBSERVATIONS. A REVISION PROCEDURE IS PLANNED, AND AVAILABLE INFORMATION SUGGESTS THAT THE PHYSICIAN WILL TRY TO REPOSITION THIS RV LEAD.
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0181 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Life Threatening | E102| 0181 |