FDA Adverse Event Injury Summary report: N

STARCLOSE SE VASCULAR CLOSURE SYSTEM

MDR report key: 1851724 · Received September 29, 2010

Report

Report Number
2953144-2010-02327
Event Type
Injury
Date Received
September 29, 2010
Date of Event
September 29, 2009
Report Date
September 29, 2009
Manufacturer
ABBOTT VASCULAR-VASCULAR SOLUTIONS
Product Code
MGB
PMA / PMN Number
P050007
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS NOT PROVIDED; THEREFORE, A DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE SE DEVICE, ACHIEVED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, AFTER CLIP DEPLOYMENT, THE DEVICE WAS DIFFICULT TO REMOVE. IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE, THE ACCESS PORTS WERE USED TO REMOVE THE DEVICE FROM THE PT'S ANATOMY. THE CLIP FROM THE STARCLOSE SE DEPLOYED IN THE INTENDED LOCATION AND ACHIEVED HEMOSTASIS. NO ADVERSE PT EFFECTS WERE REPORTED. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STARCLOSE SE VASCULAR CLOSURE SYSTEM MGB ABBOTT VASCULAR-VASCULAR SOLUTIONS NA UNK

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention