STARCLOSE SE VASCULAR CLOSURE SYSTEM
Report
- Report Number
- 2953144-2010-02299
- Event Type
- Injury
- Date Received
- September 29, 2010
- Date of Event
- August 13, 2009
- Report Date
- August 13, 2009
- Manufacturer
- ABBOTT VASCULAR - VASCULAR SOLUTIONS
- Product Code
- MGB
- PMA / PMN Number
- P050007
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVAL SUMMARY: EVAL OF THE RETURNED DEVICE FOUND THE LOCATOR WINGS WERE INITIALLY BENT DURING THUMB ADVANCER DEPLOYMENT, PREVENTING THE WINGS FROM FULLY COLLAPSING INTO THE DELIVERY TUBESET WHEN THE CLIP WAS FIRED. THE BENT LOCATOR WINGS DIRECTLY RESULTED IN DIFFICULT DEVICE REMOVAL. THE ACCESS PORTS WERE SUCCESSFULLY UTILIZED TO RELEASE THE LOCKING MECHANISM ON THE THUMB ADVANCER. ALTHOUGH THE DEVICE WAS SUCCESSFULLY REMOVED BY USING COUNTER-TRACTION AND ASSERTIVE PULL, FORCEFUL REMOVAL DIRECTLY RESULTED IN THE WINGS BEING BROKEN OFF THE DISTAL RETAINING RING. ALL WINGS REMAINED SECURE WITHIN THE LOCATOR. BASED ON THE INVESTIGATION FINDINGS, THE PROBABLE ROOT CAUSE FOR THE BENT LOCATOR WINGS THAT DIRECTLY RESULTED IN THE REPORTED DIFFICULT DEVICE REMOVAL IS TISSUE COMPACTION THAT EXERTED A DISTAL FORCE ON THE WINGS WHILE IN THE TISSUE TRACT. TISSUE TRAPPED BETWEEN THE DISTAL END OF THE TUBESET AND THE LOCATOR WINGS CAN BEND THE WINGS AND EVENTUALLY BREAK THEM DURING FORCEFUL REMOVAL. NO MFG OR QUALITY ISSUES WERE DETECTED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.
IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE SE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE IN A VESSEL GRAFT OF THE COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, AFTER CLIP DEPLOYMENT, THE DEVICE WAS DIFFICULT TO REMOVE FROM THE PT'S ARTERY. IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE, AN ATTEMPT WAS MADE TO USE THE ACCESS PORTS, HOWEVER, IT WAS UNSUCCESSFUL. THE DEVICE WAS ULTIMATELY REMOVED USING COUNTER-TRACTION AND ASSERTIVE PULL. HEMOSTASIS WAS ACHIEVED WITH THE CLIP AND LESS THAN 5 MINUTES OF MANUAL COMPRESSION. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STARCLOSE SE VASCULAR CLOSURE SYSTEM | MGB | ABBOTT VASCULAR - VASCULAR SOLUTIONS | NA | 79017-6H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |