FDA Adverse Event Injury Summary report: N

STARCLOSE VASCULAR CLOSURE SYSTEM

MDR report key: 1851676 · Received September 29, 2010

Report

Report Number
2953144-2010-02297
Event Type
Injury
Date Received
September 29, 2010
Date of Event
June 1, 2009
Report Date
August 18, 2009
Manufacturer
ABBOTT VASCULAR-VASCULAR SOLUTIONS
Product Code
MGB
PMA / PMN Number
P050007
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION OF THE RETURNED DEVICE FOUND IT WAS FULLY CLIP DEPLOYED WITH NO DETECTED DAMAGE OR ANOMALY NORMALLY ASSOCIATED WITH DIFFICULT DEVICE REMOVAL. ALL OBSERVATIONS INDICATED NORMAL DEVICE OPERATION. BASED ON THE INVESTIGATION FINDINGS, THE ROOT CAUSE FOR THE REPORTED EVENT COULD NOT BE DETERMINED. NO MANUFACTURING OR QUALITY ISSUES WERE DETECTED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE SE DEVICE ACHIEVED ARTERIOTOMY CLOSURE OF AN UNSPECIFIED VESSEL AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, AFTER DEPLOYING THE CLIP, DIFFICULTY WAS ENCOUNTERED REMOVING THE DEVICE. IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE, THE DEVICE WAS REMOVED WITH COUNTER-TRACTION AND AN ASSERTIVE PULL. HEMOSTASIS WAS ACHIEVED WITH THE CLIP. THERE WERE NO REPORTED ADVERSE PT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STARCLOSE VASCULAR CLOSURE SYSTEM MGB ABBOTT VASCULAR-VASCULAR SOLUTIONS NA 69145-6H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention