FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 1851647 · Received September 29, 2010

Report

Report Number
2953144-2010-02263
Event Type
Injury
Date Received
September 29, 2010
Date of Event
July 11, 2007
Report Date
July 14, 2007
Manufacturer
ABBOTT VASCULAR-VASCULAR SOLUTIONS
Product Code
MGB
PMA / PMN Number
P960043
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL SUMMARY: EVAL OF THE RETURNED DEVICE FOUND BOTH CUFFS WERE STILL ATTACHED TO THE LINK AND RESPECTIVE NEEDLE TIPS. THERE WAS ONLY ABOUT 3/8" OF MONOFILAMENT STILL ATTACHED TO THE NEEDLE TIP. THE REST OF THE MONOFILAMENT WAS NOT RETURNED WITH THE DEVICE. THE KNOT PUSHER SUTURE TRIMMER WAS NOT RETURNED. THERE WAS NO ABNORMAL OBSERVATION FOUND. BASED ON THE INVESTIGATION, A KNOT LOCK OCCURRED AND GOT STUCK INSIDE THE TISSUE TRACT. HOWEVER, THE DEVICE CONFORMED TO SPECIFICATION, BECAUSE THERE WERE NO ABNORMAL OBSERVATIONS WITH THE CONDITION OF THE RETURNED DEVICE THAT COULD HAVE CONTRIBUTED TO THE REPORTED COMPLAINT. BOTH CUFFS WERE STILL ATTACHED TO THE LINK AND RESPECTIVE NEEDLE TIPS. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS INVESTIGATION. BASED ON THE INVESTIGATION FINDINGS, A ROOT CAUSE FOR THE KNOT LOCK COULD NOT BE DETERMINED. NO MALFUNCTION WAS DETECTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE FEMORAL ARTERY AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, WHILE ADVANCING THE KNOT WITH THE SUTURE TRIMMER TO THE ARTERIOTOMY, THE KNOT LOCKED AND BECAME STUCK INSIDE THE TISSUE TRACT. THE DEVICE WAS REMOVED AND HEMOSTASIS WAS ACHIEVED USING MANUAL COMPRESSION. THERE WERE NO REPORTED ADVERSE PT EFFECTS. NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM MGB ABBOTT VASCULAR-VASCULAR SOLUTIONS NA 44166-6H

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention OTHER: HEPARIN