FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 1851614 · Received October 4, 2010

Report

Report Number
2124215-2010-14085
Event Type
Malfunction
Date Received
October 4, 2010
Date of Event
July 9, 2010
Report Date
July 20, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Removal / Correction Number
Z-1192-06 THRU Z-1194-06
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS UNDERGOING DETAILED LABORATORY ANALYSIS. THIS REPORT WILL BE UPDATED WHEN ANALYSIS IS COMPLETE.

Additional Manufacturer Narrative · 1

A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. ALL LOW VOLTAGE POWER SUPPLY MEASUREMENTS WERE CURRENTLY WITHIN AN APPROPRIATE RANGE; HOWEVER, FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION. THIS ISSUE IS DISCUSSED IN THE Q2 2010 PRODUCT PERFORMANCE REPORT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION AND WAS REPLACED WITH ANOTHER BOSTON SCIENTIFIC CRT-D. UPON RECEIPT, INITIAL LABORATORY ANALYSIS DETERMINED THAT THIS DEVICE DID NOT MEET EXPECTED LONGEVITY PREDICTIONS AS DESCRIBED IN DEVICE LABELING.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H170

Patients

Seq Age Sex Outcome Treatment
1 64 YR 4470| 0185| N119| 4543| H170