FDA Adverse Event Injury Summary report: N

PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1851612 · Received September 28, 2010

Report

Report Number
3004209178-2010-82922
Event Type
Injury
Date Received
September 28, 2010
Date of Event
September 11, 2010
Report Date
September 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT HE WAS TAKEN TO THE EMERGENCY ROOM FOR LOW BLOOD GLUCOSE AND RELEASED. THE BLOOD GLUCOSE READING AT TIME OF THE CALL WAS 40 MG/DL. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER STATED THAT HE HAD DONE YARD WORK FOR ABOUT TWO HOURS PRIOR TO INCIDENT. THE CUSTOMER STATED THAT HE WAS NOT ALERT WHEN HIS WIFE AND DAUGHTER FOUND HIM AND PARAMEDICS WERE CALLED. REVIEWED PRIMING TECHNIQUES AND INSULIN CONCENTRATION AND THEY WERE CORRECT. THE PROGRAMMING ON THE INSULIN PUMP WAS CORRECT. THE STATUS SCREEN UNITS MATCHED TO THE AMOUNT OF INSULIN LEFT IN THE RESERVOIR. RAN A DISPLACEMENT TEST AND THE DEVICE PASSED THE TEST. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 56 YR Hospitalization