FDA Adverse Event Injury Summary report: N

STARCLOSE SE VASCULAR CLOSURE SYSTEM

MDR report key: 1851475 · Received September 29, 2010

Report

Report Number
2953144-2010-02126
Event Type
Injury
Date Received
September 29, 2010
Date of Event
March 1, 2009
Report Date
March 2, 2009
Manufacturer
ABBOTT VASCULAR-VASCULAR SOLUTIONS
Product Code
MGB
PMA / PMN Number
P050007
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL SUMMARY: EVAL OF THE RETURNED DEVICE FOUND THREE BENT AND BROKEN VESSEL LOCATOR WINGS WITH NO OTHER DAMAGE OR ANOMALY DETECTED ON THE DEVICE, EXTERNAL OR INTERNAL. IT WAS DETERMINED THAT ALL OF THE WING MATERIAL WAS RETURNED WITH THE DEVICE BENT/BROKEN. BROKEN WINGS ARE CONSISTENT WITH DIFFICULT TO REMOVE DEVICES WHEN TISSUE COMPACTION RESULTS IN A DISTAL FORCE APPLIED TO THE LOCATOR WINGS, BENDING THEM DISTALLY, AND RESTRICTING PROPER RETRACTION INTO THE DELIVERY TUBE SET. COUNTER-TRACTION AND FORCE IS THEN APPLIED TO ACHIEVE DEVICE REMOVAL. NO OTHER ANOMALIES WERE DETECTED. BASED ON THE INVESTIGATION FINDINGS, THE ROOT CAUSE FOR THE DIFFICULT TO REMOVE CONDITION AND DAMAGED WINGS IS RELATED TO THE OPERATIONAL CONTEXT IN WHICH THE DEVICE WAS USED. NO MFG OR QUALITY ISSUES WERE DETECTED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE SE DEVICE, ACHIEVED ARTERIOTOMY CLOSURE OF A FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, AFTER CLIP DEPLOYMENT, THE DEVICE WAS DIFFICULT TO REMOVE FROM THE PT'S ARTERY. THE DEVICE WAS REMOVED USING FORCE; HEMOSTASIS WAS ACHIEVED WITH THE CLIP. IT WAS ALSO NOTED THAT ONE OF THE VESSEL LOCATOR WINGS WAS FOUND TO BE BROKEN. THERE WERE NO REPORTED ADVERSE PT EFFECTS. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STARCLOSE SE VASCULAR CLOSURE SYSTEM MGB ABBOTT VASCULAR-VASCULAR SOLUTIONS NA 70085-6H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention