FDA Adverse Event
Malfunction
Summary report: N
DURASTAR STRETCHER HYDRAULIC
MDR report key: 1851390
·
Received September 23, 2010
Report
- Report Number
- 1824206-2010-10092
- Event Type
- Malfunction
- Date Received
- September 23, 2010
- Date of Event
- August 26, 2010
- Report Date
- August 26, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECH CHECKED THE LEFT GAS CYLINDER AND IT IS FINE. TECH CHECKED THE RIGHT GAS CYLINDER AND IT WAS NOT WORKING. TECH REPLACED THE RIGHT GAS CYLINDER AND THE ISSUE RESOLVED.
Description of Event or Problem · 1
TECH ALLEGED THAT THE STRETCHER HEAD WOULD NOT GO DOWN TO THE FLAT POSITION. NO PT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DURASTAR STRETCHER HYDRAULIC | WHEELED STRETCHER | FPO | HILL-ROM, INC. | P8000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |