FDA Adverse Event Malfunction Summary report: N

LEKSELL GAMMA KNIFE PERFEXION WITH EXTEND FRAME SYSTEM

MDR report key: 1851032 · Received September 27, 2010

Report

Report Number
9612186-2010-00005
Event Type
Malfunction
Date Received
September 27, 2010
Date of Event
September 16, 2010
Report Date
September 24, 2010
Manufacturer
ELEKTA INSTRUMENT AB
Product Code
IWB
PMA / PMN Number
K092083
Removal / Correction Number
9612186-092710-016-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION RESULTS: A FAILURE OF AN EXTEND MOUTHPIECE WAS FOUND DURING A PT SETUP. THE ANTERIOR FIXATION BAR OF THE FRONTPIECE CAME LOOSE FROM THE DENTAL IMPRESSION TRAY. THE JOINT BETWEEN THE ANTERIOR FIXATION BAR AND THE DENTAL IMPRESSION TRAY IS GLUED USING CYANOACRYLATE GLUE. THE BOND GAP HAS BEEN TOO LARGE IN SOME MOUTHPIECES, RESULTING IN A TOO SMALL CONTACT SURFACE BETWEEN THE TWO PARTS AND AS A CONSEQUENCE A TOO WEAK JOINT. THE RESULT IS THAT THE GLUE LINE CAN BREAK WHEN THE MOUTHPIECE IS TWISTED, WHICH LEADS TO EITHER A PLAY BETWEEN THE ANTERIOR FIXATION BAR AND THE DENTAL IMPRESSION TRAY, OR THAT THE DENTAL IMPRESSION TRAY COMES LOOSE. IT IS NOT POSSIBLE TO DETERMINE IF THE GLUE LINES ON A MOUTHPIECE ARE BAD, NEITHER BY VISUAL INSPECTION NOR BY PULLING OR BENDING THE EXTEND MOUTHPIECE. CORRECTIVE ACTION: ELEKTA INSTRUMENTS (B)(4) ISSUED AN IMPORTANT NOTICE A342 "POTENTIAL FAILURE OF THE EXTEND MOUTHPIECE" TO AFFECTED USERS INSTRUCTING THEM TO RETURN THE OLD MOUTHPIECES TO ELEKTA INSTRUMENTS (B)(4) WHICH WILL BE REPLACED WITH NEW MOUTHPIECES. THE CORRECTIVE ACTIONS HAVE BEEN IDENTIFIED WITHIN THE (B)(4).

Description of Event or Problem · 1

DURING SETUP OF PT AT CT, IT WAS DIFFICULT TO ACHIEVE GOOD RESULTS AT RCT MEASUREMENT. THE ROD TURNED OUT TO HAVE LOOSENED FROM THE MOUTHPIECE. ALL MOUTHPIECES AT CUSTOMERS WERE THEN CHECKED BY HAND FORCE PRIOR TO THE NEXT SETUP TRIAL. AFTER CT PRIOR TO TREATMENT, THE SAME THING HAPPENED AGAIN. SIZE OF THE MOUTHPIECE WAS XL P2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LEKSELL GAMMA KNIFE PERFEXION WITH EXTEND FRAME SYSTEM SYSTEM, RADIATION THERAPY, RADIONUCLIDE IWB ELEKTA INSTRUMENT AB

Patients

Seq Age Sex Outcome Treatment
1