LEKSELL GAMMA KNIFE PERFEXION WITH EXTEND FRAME SYSTEM
Report
- Report Number
- 9612186-2010-00005
- Event Type
- Malfunction
- Date Received
- September 27, 2010
- Date of Event
- September 16, 2010
- Report Date
- September 24, 2010
- Manufacturer
- ELEKTA INSTRUMENT AB
- Product Code
- IWB
- PMA / PMN Number
- K092083
- Removal / Correction Number
- 9612186-092710-016-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- UNKNOWN
Narratives
INVESTIGATION RESULTS: A FAILURE OF AN EXTEND MOUTHPIECE WAS FOUND DURING A PT SETUP. THE ANTERIOR FIXATION BAR OF THE FRONTPIECE CAME LOOSE FROM THE DENTAL IMPRESSION TRAY. THE JOINT BETWEEN THE ANTERIOR FIXATION BAR AND THE DENTAL IMPRESSION TRAY IS GLUED USING CYANOACRYLATE GLUE. THE BOND GAP HAS BEEN TOO LARGE IN SOME MOUTHPIECES, RESULTING IN A TOO SMALL CONTACT SURFACE BETWEEN THE TWO PARTS AND AS A CONSEQUENCE A TOO WEAK JOINT. THE RESULT IS THAT THE GLUE LINE CAN BREAK WHEN THE MOUTHPIECE IS TWISTED, WHICH LEADS TO EITHER A PLAY BETWEEN THE ANTERIOR FIXATION BAR AND THE DENTAL IMPRESSION TRAY, OR THAT THE DENTAL IMPRESSION TRAY COMES LOOSE. IT IS NOT POSSIBLE TO DETERMINE IF THE GLUE LINES ON A MOUTHPIECE ARE BAD, NEITHER BY VISUAL INSPECTION NOR BY PULLING OR BENDING THE EXTEND MOUTHPIECE. CORRECTIVE ACTION: ELEKTA INSTRUMENTS (B)(4) ISSUED AN IMPORTANT NOTICE A342 "POTENTIAL FAILURE OF THE EXTEND MOUTHPIECE" TO AFFECTED USERS INSTRUCTING THEM TO RETURN THE OLD MOUTHPIECES TO ELEKTA INSTRUMENTS (B)(4) WHICH WILL BE REPLACED WITH NEW MOUTHPIECES. THE CORRECTIVE ACTIONS HAVE BEEN IDENTIFIED WITHIN THE (B)(4).
DURING SETUP OF PT AT CT, IT WAS DIFFICULT TO ACHIEVE GOOD RESULTS AT RCT MEASUREMENT. THE ROD TURNED OUT TO HAVE LOOSENED FROM THE MOUTHPIECE. ALL MOUTHPIECES AT CUSTOMERS WERE THEN CHECKED BY HAND FORCE PRIOR TO THE NEXT SETUP TRIAL. AFTER CT PRIOR TO TREATMENT, THE SAME THING HAPPENED AGAIN. SIZE OF THE MOUTHPIECE WAS XL P2.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LEKSELL GAMMA KNIFE PERFEXION WITH EXTEND FRAME SYSTEM | SYSTEM, RADIATION THERAPY, RADIONUCLIDE | IWB | ELEKTA INSTRUMENT AB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |