FDA Adverse Event
Malfunction
Summary report: N
REPAIR TPS SAG SAW
MDR report key: 1851030
·
Received September 27, 2010
Report
- Report Number
- 1811755-2010-01252
- Event Type
- Malfunction
- Date Received
- September 27, 2010
- Date of Event
- August 31, 2010
- Report Date
- August 31, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE HAS BEEN RECEIVED AT THE MANUFACTURER FOR INVESTIGATION. AN EVAL WAS CONDUCTED AND THE COMPLAINT WAS CONFIRMED. ACCORDING TO THE INVESTIGATION DETAILS, DEBRIS WAS FOUND IN THE ROTOR. THE BEARINGS, MOTOR, AND PRESS PLUG WERE REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE OVERHEATED DURING TESTING. THIS DID NOT OCCUR DURING ANY SURGICAL OR MEDICAL PROCEDURE, SO THERE WAS NO PT INVOLVEMENT. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REPAIR TPS SAG SAW | BONE CUTTING INSTRUMENT AND ACCESSORIES | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |