FDA Adverse Event
Malfunction
Summary report: N
MICRO DRILL MEDIUM STRAIGHT ATTACHMENT
MDR report key: 1850999
·
Received September 23, 2010
Report
- Report Number
- 1811755-2010-01243
- Event Type
- Malfunction
- Date Received
- September 23, 2010
- Date of Event
- September 7, 2010
- Report Date
- September 7, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE ATTACHMENT WAS RECEIVED AT THE MANUFACTURER FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS CONFIRMED. THE DEVICE WAS DISASSEMBLED, AND CORROSION AND DEBRIS WERE FOUND INSIDE. BASED ON THE INVESTIGATION DETAILS, THE FAILURE OF THE ATTACHMENT WAS MOST LIKELY CAUSED BY CORROSION AND DEBRIS PRESENT ON ASSEMBLY COMPONENTS, INCLUDING BEARINGS. THIS WAS LIKELY THE RESULT OF INADEQUATE CLEANING BY NOT FOLLOWING THE INSTRUCTIONS FOR USE. THE ATTACHMENT WAS SCRAPPED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ATTACHMENT BEGAN OVERHEATING DURING A SURGICAL PROCEDURE. THE CASE WAS CONTINUED WITH ANOTHER DEVICE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICRO DRILL MEDIUM STRAIGHT ATTACHMENT | SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC | ERL | STRYKER INSTRUMENTS KALAMAZOO | 08350 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |