FDA Adverse Event Malfunction Summary report: N

MICRO DRILL MEDIUM STRAIGHT ATTACHMENT

MDR report key: 1850999 · Received September 23, 2010

Report

Report Number
1811755-2010-01243
Event Type
Malfunction
Date Received
September 23, 2010
Date of Event
September 7, 2010
Report Date
September 7, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE ATTACHMENT WAS RECEIVED AT THE MANUFACTURER FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS CONFIRMED. THE DEVICE WAS DISASSEMBLED, AND CORROSION AND DEBRIS WERE FOUND INSIDE. BASED ON THE INVESTIGATION DETAILS, THE FAILURE OF THE ATTACHMENT WAS MOST LIKELY CAUSED BY CORROSION AND DEBRIS PRESENT ON ASSEMBLY COMPONENTS, INCLUDING BEARINGS. THIS WAS LIKELY THE RESULT OF INADEQUATE CLEANING BY NOT FOLLOWING THE INSTRUCTIONS FOR USE. THE ATTACHMENT WAS SCRAPPED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATTACHMENT BEGAN OVERHEATING DURING A SURGICAL PROCEDURE. THE CASE WAS CONTINUED WITH ANOTHER DEVICE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO DRILL MEDIUM STRAIGHT ATTACHMENT SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC ERL STRYKER INSTRUMENTS KALAMAZOO 08350

Patients

Seq Age Sex Outcome Treatment
1 UNK