FDA Adverse Event Injury Summary report: N

STYLE 15 SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 18508763 · Received January 12, 2024

Report

Report Number
9617229-2024-00864
Event Type
Injury
Date Received
January 12, 2024
Report Date
March 7, 2024
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE EVENT OF INFECTION (LATE ONSET) IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: BASED ON THE DEVICE ANALYSIS GRID, THE ASSESSMENTS OF THE COMPLAINT ARE: ¿ RUPTURE: OBSERVED BROKEN DEVICE ASSESSED AS SURGICAL DAMAGE. ¿ INFECTION (LATE ONSET: UNABLE TO OBSERVE SINCE IT IS NOT RELATED TO THE DEVICE. ¿ ERYTHEMA: UNABLE TO OBSERVE SINCE IT IS NOT RELATED TO THE DEVICE. ¿ EDEMA: UNABLE TO OBSERVE SINCE IT IS NOT RELATED TO THE DEVICE. NO FURTHER ACTIONS ARE REQUIRED AS NO MANUFACTURING ISSUES ARE OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INVESTIGATION: WITH THE INFORMATION COLLECTED DURING THE INVESTIGATION, THERE IS ENOUGH EVIDENCE TO SUPPORT THAT DEVICE LOT NUMBER 2011987 WAS MANUFACTURED UNDER CONTROLLED CONDITIONS IN ACCORDANCE WITH ABBVIE¿S PROCEDURES AND WERE NO DEFECTS/PROBLEMS/ISSUES FOUND IN THE DOCUMENTS RELATED TO THE REPORTED EVENT. SEAL STRENGTH WAS SUCCESSFULLY COMPLETED AND RESULTS WERE ACCEPTABLE. ENVIRONMENTAL MONITORING RESULTS WERE FOUND TO BE ACCEPTABLE, AND THEY CONFIRMED THAT THERE WAS NOT AN ADVERSE IMPACT ON ENVIRONMENTAL CONDITIONS, DEVICE QUALITY OR PERFORMANCE. THE STERILIZATION RUN (B)(6) IS NOT RELATED TO ANY ADDITIONAL COMPLAINT INFECTION RECORD REPORTED AS OF TODAY. DURING THE TREND REVIEW OF ALL INFECTION COMPLAINTS FOR GEL FOR THE PERIOD OF MAR 2022 THROUGH FEB 2024, WERE NOTED ONE OUTLIER IN JUN 2022. THE ADDITIONAL ANALYSIS PERFORMED SHOWS ALL THAT RESULTS MET THE ACCEPTANCE CRITERIA AND NO ISSUES, DEVIATIONS OR NON-CONFORMANCES WERE FOUND WHICH COULD BE ASSOCIATED TO THE INFECTION EVENT, SO, NO ADDITIONAL ACTIONS ARE DEEMED REQUIRED AT THIS TIME. THE ISSUE WITH INFECTION WILL CONTINUE TO BE MONITORED AND CORRECTIVE ACTION TAKEN IN THE FUTURE IF DEEMED APPROPRIATE. GIVEN THE FACT THAT THE CAUSE OF THE INFECTION CANNOT BE SPECIFICALLY ASSOCIATED TO THE MANUFACTURING PROCESS, AND THERE IS NOT AN ADVERSE TREND FOR THIS TYPE OF EVENT, AND NO ER/ NCR(S) OR TEMPORARY CHANGES WERE IDENTIFIED DURING THE INVESTIGATION ASSOCIATED TO THIS LOT AND THE COMPLAINT, NO CORRECTIVE ACTION IS DEEMED NECESSARY AT THIS TIME.

Additional Manufacturer Narrative · 0

FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.REASON FOR REOPERATION: RUPTURE.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL ADDITIONALLY REPORTED "INFECTED BREAST, DRAINAGE OF LEFT BREAST ABSCESS, ERYTHEMA, AND SWELLING." DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1846683 STYLE 15 SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2011987

Patients

Seq Age Sex Outcome Treatment
1 NA Female Required Intervention