FDA Adverse Event Malfunction Summary report: N

BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER

MDR report key: 18508673 · Received January 12, 2024

Report

Report Number
1710034-2023-01533
Event Type
Malfunction
Date Received
January 12, 2024
Date of Event
November 9, 2023
Report Date
June 24, 2024
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825332
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

CORRECTION: F CODE CORRECTED. INVESTIGATION RESULTS: NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 382533 AND LOT NUMBER 3251082. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. BASED ON THE AVAILABLE INFORMATION, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN UNSPECIFIED AMOUNT OF BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER NEEDLES WOULD NOT RETRACT WHEN THE SAFETY BUTTON WAS PRESSED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THEY HAD TROUBLE RETRACTING THE NEEDLE."

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2058476 BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 3251082 00382903825332

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown