INFANT NASAL CANNULA OXYGEN THERAPY

MDR report key: 1850821 · Received October 3, 2010

Report

Report Number
9611451-2010-00609
Event Type
Malfunction
Date Received
October 3, 2010
Date of Event
September 7, 2010
Report Date
September 7, 2010
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
CAT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE RETURNED CANNULA WAS VISUALLY INSPECTED FOR DAMAGE OR ABNORMALITIES. RESULTS: THE CANNULA TUBING WAS FOUND TO BE DISCONNECTED FROM THE NASAL INTERFACE, CONFIRMING THE REPORTED FAULT. THERE WAS NO RESIDUE ON THE TUBING AND NO DAMAGE TO THE SURFACE OF THE TUBING. A LOT CHECK COULD NOT BE CARRIED OUT AS NO LOT NUMBER WAS PROVIDED. CONCLUSION: THERE WAS NO RESIDUE ON THE TUBING. THIS SUGGESTS THAT THERE WAS NO BONDING AGENT OR AN INSUFFICIENT AMOUNT OF BONDING AGENT APPLIED TO THE CANNULA DURING MANUFACTURING. AS THERE WAS NO DAMAGE TO THE SURFACE OF THE TUBING, IT IS LIKELY THAT ANY BONDING MATERIAL APPLIED HAD NOT BONDED PROPERLY, AS A BOND BROKEN BY FORCE WILL DAMAGE THE TUBE SURFACE. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT THE TUBING ON A BC2745 INFANT NASAL CANNULA "KEPT DETACHING FROM THE CANNULA ITSELF." NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT NASAL CANNULA OXYGEN THERAPY CAT CAT FISHER & PAYKEL HEALTHCARE LIMITED BC2745-20 NOT PROVIDED

Patients

Seq Age Sex Outcome Treatment
1