FDA Adverse Event
Summary report: N
UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 1850635
·
Received October 1, 2010
Report
- Report Number
- 2050012-2010-00922
- Date Received
- October 1, 2010
- Date of Event
- August 31, 2010
- Report Date
- October 1, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER WAS SENT A REPLACEMENT.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO A BROKEN CARTRIDGE PACKAGE. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | N/A | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |