FDA Adverse Event
Malfunction
Summary report: N
REMANUFACTURED
MDR report key: 1850394
·
Received September 20, 2010
Report
- Report Number
- 1824206-2010-10033
- Event Type
- Malfunction
- Date Received
- September 20, 2010
- Date of Event
- August 25, 2010
- Report Date
- August 25, 2010
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN FOUND A WAX/DIRT BUILD UP ON THE CASTERS' WHEEL. TECHNICIAN WAS UNABLE TO CLEAN THE BUILD UP OFF OF THE CASTER WHEEL. TECHNICIAN ALSO FOUND SOME CRACKING ON THE RUBBER OF THE CASTER WHEEL. TECHNICIAN REPLACED THE BRAKE STEER CASTER AND BRAKE CASTERS TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
TECHNICIAN ALLEGED THAT THE BED WAS SLIDING WHILE IN BRAKE MODE.