FDA Adverse Event Malfunction Summary report: N

REMANUFACTURED

MDR report key: 1850394 · Received September 20, 2010

Report

Report Number
1824206-2010-10033
Event Type
Malfunction
Date Received
September 20, 2010
Date of Event
August 25, 2010
Report Date
August 25, 2010
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND A WAX/DIRT BUILD UP ON THE CASTERS' WHEEL. TECHNICIAN WAS UNABLE TO CLEAN THE BUILD UP OFF OF THE CASTER WHEEL. TECHNICIAN ALSO FOUND SOME CRACKING ON THE RUBBER OF THE CASTER WHEEL. TECHNICIAN REPLACED THE BRAKE STEER CASTER AND BRAKE CASTERS TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

TECHNICIAN ALLEGED THAT THE BED WAS SLIDING WHILE IN BRAKE MODE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REMANUFACTURED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. P8400

Patients

Seq Age Sex Outcome Treatment
1