Description of Event or Problem · 1

THE TECHNICIAN ALLEGED THE NURSE CALL SYSTEM WAS NOT WORKING CORRECTLY. WHEN A CALL WAS PLACED USING BEDRAIL SWITCHES, MASTER STATION RESPONSE NOT HEARD IN EXPECTED LOCATION. THE TECHNICIAN FOUND LOOSE PINS IN THE COMMUNICATION CABLE 37 PIN CONNECTOR. HE INSTALLED A CABLE SAVER, REPAIRED THE PINS AND TESTED FUNCTIONS TO RESOLVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANCE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. P1155F54