FDA Adverse Event
Malfunction
Summary report: N
ADVANCE
MDR report key: 1850389
·
Received September 20, 2010
Report
- Report Number
- 1824206-2010-10015
- Event Type
- Malfunction
- Date Received
- September 20, 2010
- Date of Event
- August 27, 2010
- Report Date
- August 27, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE TECHNICIAN ALLEGED THE NURSE CALL SYSTEM WAS NOT WORKING CORRECTLY. WHEN A CALL WAS PLACED USING BEDRAIL SWITCHES, MASTER STATION RESPONSE NOT HEARD IN EXPECTED LOCATION. THE TECHNICIAN FOUND LOOSE PINS IN THE COMMUNICATION CABLE 37 PIN CONNECTOR. HE INSTALLED A CABLE SAVER, REPAIRED THE PINS AND TESTED FUNCTIONS TO RESOLVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVANCE | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | P1155F54 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |