FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1850380
·
Received September 20, 2010
Report
- Report Number
- 2531779-2010-01294
- Event Type
- Malfunction
- Date Received
- September 20, 2010
- Date of Event
- August 19, 2010
- Report Date
- August 19, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED A DISLODGED DISPLAY SCREEN. LOSS OF CARTRIDGE DETECTION WAS FOUND DURING THE REVIEW OF THE PUMP HISTORY, WHICH WAS NOT DUPLICATED DURING INVESTIGATION.
Description of Event or Problem · 1
A REVIEW OF THE PUMP HISTORY REVEALED LOSS OF CARTRIDGE DETECTION. DURING EVAL, THE DISPLAY SCREEN WAS FOUND TO BE DISLODGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |