FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1850378 · Received September 20, 2010

Report

Report Number
2531779-2010-01295
Event Type
Malfunction
Date Received
September 20, 2010
Date of Event
August 19, 2010
Report Date
August 19, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH COULD NOT BE DUPLICATED DURING TESTING. EVAL ALSO REVEALED A MISALIGNED DISPLAY SCREEN.

Description of Event or Problem · 1

A REVIEW OF THE PUMP HISTORY REVEALED LOSS OF CARTRIDGE DETECTION. DURING EVAL, THE DISPLAY SCREEN WAS FOUND TO BE DISLODGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 8 YR