LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 1850280 · Received October 1, 2010

Report

Report Number
3005075853-2010-05631
Event Type
Malfunction
Date Received
October 1, 2010
Date of Event
August 1, 2010
Report Date
September 7, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). MALFORMED CLIP. THE ANALYSIS RESULTS OF THE (B)(4) DEVICE FOUND THAT IT WAS RECEIVED WITH JAWS IN THE CLOSED POSITION. THE TRIGGER WAS MANUALLY ASSISTED IN ORDER TO OPEN THE JAWS WITHOUT ANY DIFFICULTIES NOTED; ONE PEAR SHAPED CLIP WAS RELEASED. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. IT SHOULD BE NOTED THAT AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE OPENING ISSUES. THE REMAINING CLIPS WERE CONFORMING ACCORDING TO OUR MANUFACTURING SPECIFICATIONS. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE INDICATOR WAS NOT COMPLETELY VISIBLE. THIS FINDING IS NOT RELATED TO THE INCIDENT REPORTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE DEVICE WAS FORMING MALFORMED CLIPS. THEY COMPLETED THE PROCEDURE WITH A NEW LIKE DEVICE. THERE WAS NO PATIENT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM ENDO CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA G4TE8H

Patients

Seq Age Sex Outcome Treatment
1