FDA Adverse Event Injury Summary report: N

VITALITY 2

MDR report key: 1850254 · Received October 1, 2010

Report

Report Number
2124215-2010-14748
Event Type
Injury
Date Received
October 1, 2010
Date of Event
July 5, 2010
Report Date
July 5, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE. THIS DEVICE IS INCLUDED IN THE MID-LIFE DISPLAY OF REPLACEMENT INDICATORS (NOVEMBER 27, 2007) ADVISORY POPULATION. THIS ISSUE IS DISCUSSED IN THE Q2 2010 PRODUCT PERFORMANCE REPORT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DISPLAYED A BATTERY STATUS OF END OF LIFE (EOL). CURRENTLY, THE CHARGE TIME WAS 34.3 SECONDS. AT THE PRIOR FOLLOW-UP VISIT SIX MONTHS AGO, THE CHARGE TIME WAS 15.6 SECONDS WITH A BATTERY STATUS OF MIDDLE OF LIFE 2 (MOL2). THERE WAS A CONCERN THAT THE BATTERY DEPLETED EARLIER THAN EXPECTED. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED. THIS DEVICE WAS EXPLANTED AND RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T167

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening| R