FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-522NAS PRDGM INS V2.2 SK EN
MDR report key: 1850165
·
Received September 22, 2010
Report
- Report Number
- 3004209178-2010-82891
- Event Type
- Malfunction
- Date Received
- September 22, 2010
- Date of Event
- August 31, 2010
- Report Date
- August 31, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AS A RESULT OF A CORRODED BATTERY TUBE AND BATTERY CAP. FURTHER TESTING WAS NOT PERFORMED DUE TO THE BLANK DISPLAY.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP SHUT OFF INADVERTENTLY. THE CUSTOMER STATED THAT HE TRIED TO BOLUS, BUT THE DEVICE DID GO FURTHER THAN 0.5 UNITS AND STARTED OVER AGAIN. THE CUSTOMER ALSO STATED THAT THE NUMBERS ON THE DEVICE DISAPPEARED. TROUBLESHOOTING WAS PERFORMED. THE BOLUS HISTORY REVEALED THAT A DELIVERY OF 1.5 UNITS WAS ENTERED AND IT STOPPED AT 0.8 UNITS. THE ALARM HISTORY REVEALED SEVERAL LOW RESERVOIR ALARMS. RAN A 1.0 UNIT TEST AND THE INSULIN EXITED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-522NAS PRDGM INS V2.2 SK EN | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-522NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR |