FDA Adverse Event Malfunction Summary report: N

PUMP MMT-522NAS PRDGM INS V2.2 SK EN

MDR report key: 1850165 · Received September 22, 2010

Report

Report Number
3004209178-2010-82891
Event Type
Malfunction
Date Received
September 22, 2010
Date of Event
August 31, 2010
Report Date
August 31, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AS A RESULT OF A CORRODED BATTERY TUBE AND BATTERY CAP. FURTHER TESTING WAS NOT PERFORMED DUE TO THE BLANK DISPLAY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP SHUT OFF INADVERTENTLY. THE CUSTOMER STATED THAT HE TRIED TO BOLUS, BUT THE DEVICE DID GO FURTHER THAN 0.5 UNITS AND STARTED OVER AGAIN. THE CUSTOMER ALSO STATED THAT THE NUMBERS ON THE DEVICE DISAPPEARED. TROUBLESHOOTING WAS PERFORMED. THE BOLUS HISTORY REVEALED THAT A DELIVERY OF 1.5 UNITS WAS ENTERED AND IT STOPPED AT 0.8 UNITS. THE ALARM HISTORY REVEALED SEVERAL LOW RESERVOIR ALARMS. RAN A 1.0 UNIT TEST AND THE INSULIN EXITED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-522NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-522NAS

Patients

Seq Age Sex Outcome Treatment
1 53 YR