FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1850133
·
Received September 21, 2010
Report
- Report Number
- 2531779-2010-01308
- Event Type
- Malfunction
- Date Received
- September 21, 2010
- Date of Event
- August 23, 2010
- Report Date
- August 23, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS NOT RETURNED TO ANIMAS FOR EVAL. IF THE DEVICE IS RETURNED, AN EVAL WILL BE CONDUCTED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
Description of Event or Problem · 1
THE DISTRIBUTOR REPORTED THAT THE PUMP DISPENSED INSULIN FROM THE CARTRIDGE DURING THE LOAD STEP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |