FDA Adverse Event Malfunction Summary report: N

TOTALCARE-TREAT SURFACE&SCALE

MDR report key: 1850128 · Received September 28, 2010

Report

Report Number
1824206-2010-10164
Event Type
Malfunction
Date Received
September 28, 2010
Date of Event
September 2, 2010
Report Date
September 2, 2010
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND THAT THE VALVE GUIDE TUBE ASSEMBLY WAS DEFECTIVE. TECHNICIAN REPLACED THE VALVE GUIDE TUBE TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

ACCOUNT STATED THAT THEY HAD A BED THAT WAS DRIFTING DOWN VERY SLOWLY. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE-TREAT SURFACE&SCALE A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. P1900

Patients

Seq Age Sex Outcome Treatment
1