Additional Manufacturer Narrative · 1
THE TECHNICIAN REPLACED THE BRAKE DETENT TO REPAIR THE BED. MOD 424 IS A FIELD CORRECTIVE ACTION WHICH REPLACES THE BRAKE DETENT MECHANISM, BRAKE/STEER PEDALS AND ADJUSTS ALL FOUR CASTERS OF THE AFFINITY 4 BIRTHING BED. THIS FAILURE OCCURRED POST MODIFICATION.
Description of Event or Problem · 1
INFO RECEIVED INDICATES, THE BRAKE PEDAL WOULD RELEASE FROM THE LOCKED POSITION IF THE BED IS BUMPED.