FDA Adverse Event Malfunction Summary report: N

AFFINITY 4

MDR report key: 1850095 · Received September 28, 2010

Report

Report Number
1824206-2010-10178
Event Type
Malfunction
Date Received
September 28, 2010
Date of Event
September 1, 2010
Report Date
September 1, 2010
Manufacturer
HILL-ROM INC.
Product Code
HDD
Removal / Correction Number
Z-0548-2010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE BRAKE DETENT TO REPAIR THE BED. MOD 424 IS A FIELD CORRECTIVE ACTION WHICH REPLACES THE BRAKE DETENT MECHANISM, BRAKE/STEER PEDALS AND ADJUSTS ALL FOUR CASTERS OF THE AFFINITY 4 BIRTHING BED. THIS FAILURE OCCURRED POST MODIFICATION.

Description of Event or Problem · 1

INFO RECEIVED INDICATES, THE BRAKE PEDAL WOULD RELEASE FROM THE LOCKED POSITION IF THE BED IS BUMPED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 4 BIRTHING BED HDD HILL-ROM INC. 3700B

Patients

Seq Age Sex Outcome Treatment
1