FDA Adverse Event Malfunction Summary report: N

PUMP MMT-1712KL 640G V4.10 BK SF MG

MDR report key: 18492397 · Received January 11, 2024

Report

Report Number
2032227-2024-105506
Event Type
Malfunction
Date Received
January 11, 2024
Date of Event
December 21, 2023
Report Date
July 9, 2024
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
000000763000317195
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

NO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM¿NOTED DURING BOOT UP. THE PUMP PASSED THE FUNCTIONAL TESTS, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND DAT AT 0.08745 INCHES. NO UNEXPECTED CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM NOTED DURING THE TESTING. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. SUCCESSFULLY DOWNLOADED COMLINK3 FILE. THERE WERE NO FATAL CRITICAL ALARMS, OR THREE CONSECUTIVE CRITICAL HANDLING ALARMS FOUND IN THE FORMATTED HISTORY FILE. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM WAS DUE TO THE RF TEST FAILURE IN THE BOOTLOADER (CODE 0X00000045) ACCORDING IN THE COMLINK3 FILE, SUSPECTED ON HW. PLEASE SEE BELOW FOR PUMP ERRORS/ALARMS NOTED 2 DAYS PRIOR TO THE EVENT DATE 21-DEC-2023 IN THE FORMATTED HISTORY FILE. SENSOR EXPIRED ALERT (794) WAS FOUND ON: 12/20/2023 17:33:14.000 THE PUMP WAS PROGRAMMED WITH A TEST GUARDIAN LINK (3) TRANSMITTER AND A 240 MG/DL GLUCOSE SENSOR SIMULATOR. THE PUMP CONNECTED SUCCESSFULLY TO THE TRANSMITTER AND DISPLAYED ¿TRANSMITTER CONNECTION SUCCESSFUL¿. THE PUMP COMMUNICATED PROPERLY WITH GLUCOSE SENSOR SIMULATOR AND DISPLAYED THE CALIBRATE YOUR SENSOR ALARM PROPERLY AFTER COMPLETION OF THE WARM UP. THE PUMP CALIBRATED AND DISPLAYED THE PROGRAMMED VALUE OF 239 MG/DL PROPERLY ON THE DISPLAY GRAPH. NO SENSOR EXPIRED ALERT OR UNEXPECTED SENSOR ERRORS OR ANOMALIES WERE NOTED DURING TESTING. PUMP ERROR 63 ALARM (VARIABLEINFO = 15) WAS FOUND ON: 12/20/2023 18:23:18.000 PUMP ERROR 68 ALARM WAS FOUND ON: 12/20/2023 18:23:20.000 PUMP ERROR 49 ALARM WAS FOUND ON: 12/20/2023 18:23:20.000 12/20/2023 18:25:19.000 PUMP ERROR 23 ALARM WAS FOUND ON: 12/20/2023 18:25:39.000 PUMP ERROR 63 ALARM (VARIABLEINFO = 15) WAS CONFIRMED ACCORDING IN THE FORMATTED HISTORY FILE, SUSPECTED ON HW. PUMP ERROR 68 ALARM, PUMP ERROR 49 ALARM AND PUMP ERROR 23 ALARM WERE EXPECTED SINCE THE PUMP RESTARTED DUE TO PUMP ERROR 63 ALARM (VARIABLEINFO = 15). PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND¿NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE PCBA 1, PCBA 2, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A SCRATCHED CASE. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM WAS CONFIRMED, SUSPECTED ON HW. PUMP ERROR 63 ALARM (VARIABLEINFO = 15) WAS CONFIRMED ACCORDING IN THE FORMATTED HISTORY FILE, SUSPECTED ON HW. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED OUTSIDE THE UNITED STATES. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER RECEIVED A CRITICAL PUMP ERROR/OPEN BOOK IMAGE. TROUBLESHOOTING WAS PERFORMED, EXPLAINED THE PUMP PERFORMS SAFETY CHECKS AND AN ERROR WAS FOUND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WAS INSTRUCTED TO REVERT TO THE BACKUP PLAN PER HEALTH CARE PROFESSIONAL INSTRUCTION AND THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
706219 PUMP MMT-1712KL 640G V4.10 BK SF MG PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1712KL NG3223499H 000000763000317195

Patients

Seq Age Sex Outcome Treatment
1 40 YR Unknown